EasyManuals Logo

Agilent Technologies 6500 Series User Manual

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 77
Maintenance 2
To clean the spray chamber weekly for the ESI with Agilent Jet Stream
15 Wipe the area around the spray shield.
16 Close the spray chamber.
17 Reinstall the electrospray nebulizer.
CAUTION
Do not spray directly toward the tip of the capillary. This can cause pressure surges in
the vacuum system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals