84 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To adjust the nebulizer needle for the ESI with Agilent Jet Stream
Figure 33 Adjusting the needle holder
5 Tighten the T6 screw. Make sure this does not change the position of the
needle.
6 Remove the nebulizer from the adjustment fixture and reinstall it in the
electrospray spray chamber.
7 Be very careful not to hit the tip of the nebulizer against anything. Any
damage will have a large, negative effect on system performance.
Record this procedure in the Maintenance Logbook.