Firmware Upgrades Firmware Upgrade Process
292 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
6 Select the com port where the serial cable is connected. Choose the maximum
baud rate. Click on OK.
7 Turn on the 8166A and wait for the first beep.
Now you must be quick:
Press and hold the Preset hardkey.
Still holding Preset, press and hold the Aux hardkey.
Still holding Preset and Aux, press and hold the Appl hardkey.
TIP Use three fingers and place them near the keys in advance!
If you were successful, you here a double beep and the display shows that the
instrument is in download mode. Otherwise, turn off the power and try again.
8 Click on Program and wait until the download completes. The instrument will
reboot with the new firmware.