Chapter 3 Measurement
3-76
AF3 – AF3 Tone Frequency
Summary
Sets the Tone frequency of AF3 signal source.
This setting is enabled when the MS2830A-018/118 is installed.
Setting range
20.0 to 40000.0 Hz
AF3 – AF3 Tone Deviation (FM)
Summary
Sets the Deviation of AF3 signals.
Displayed when the MS2830A-018/118 is installed and Modulation
is set to FM.
Setting range
0.0 to 100000.0 Hz
AF3 – AF3 Tone Radian (
M)
Summary
Sets the Deviation of AF3 signals.
Displayed when the MS2830A-018/118 is installed and Modulation
is set to
M.
Setting range
0.00 to 50.00 rad
However, the setting range is limited to within the range
which meets the following formula.
(AF3 Tone Frequency
AF3 Tone Radian) < 100k
AF3 – AF3 Tone Depth (AM)
Summary
Sets the AM modulation index of AF3 signal.
Displayed when the MS2830A-018/118 is installed and Modulation
is set to AM.
Setting range
0 to 100%
Auto Set
Summary
Selects whether to allow value change in the dialog box without
determining by the Set key .