3.7 RX Measurement Mode
3-77
3
Measurement
■
Options
On Enables setting change by the knob or up/down
keys without determining by the Set key on the
AF Tone setting screens.
Off Value changed by the knob or up/down keys will
not be set without determining by the Set key .
Digital Code Squelch
■
Summary
Selects the signal output of DCS signal.
Enabled when Modulation is set to FM.
■
Options
Off Does not generate signals.
On Generates DCS signals.
Digital Code Squelch Data
■
Summary
Sets DCS Code.
■
Setting range
0 to 777 (three-digit octal notation)
Or
000…0 to 111…1 (23-digit binary notation)
3.8 “Digital Code Squelch Data Setting”
Digital Code Squelch Deviation
■
Summary
Sets the Deviation of DCS signals.
■
Setting range
0.0 to 100000.0 Hz
Digital Code Squelch Polarity
■
Summary
Sets polarity of DCS signal.
■
Setting range
Normal Polarity of DCS Code is not inverted.
Inverted Polarity of DCS Code is inverted.