EasyManua.ls Logo

Anritsu MX269018A - Page 119

Anritsu MX269018A
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.7 RX Measurement Mode
3-77
3
Measurement
Options
On Enables setting change by the knob or up/down
keys without determining by the Set key on the
AF Tone setting screens.
Off Value changed by the knob or up/down keys will
not be set without determining by the Set key .
Digital Code Squelch
Summary
Selects the signal output of DCS signal.
Enabled when Modulation is set to FM.
Options
Off Does not generate signals.
On Generates DCS signals.
Digital Code Squelch Data
Summary
Sets DCS Code.
Setting range
0 to 777 (three-digit octal notation)
Or
000…0 to 111…1 (23-digit binary notation)
3.8 “Digital Code Squelch Data Setting”
Digital Code Squelch Deviation
Summary
Sets the Deviation of DCS signals.
Setting range
0.0 to 100000.0 Hz
Digital Code Squelch Polarity
Summary
Sets polarity of DCS signal.
Setting range
Normal Polarity of DCS Code is not inverted.
Inverted Polarity of DCS Code is inverted.

Table of Contents

Related product manuals