Service Manual i2000SR interface module
625798100.APS.5.doc Page 55 of 212
Figure 32 – Electrical panel top connectors
No. Description
A0 & A1 Track module antennas A0 and A1
A2 to A8 i2000SR IM antennas from A2 to A8
J1 i2000SR IM three Motors
(Lane 1 & 2 motor, Lane 3 & 4 motor, and Disk motor)
J2 Instrument Indicator light (IM Status pole lamps)
J3 Track CAN bus power supply 24 V DC
J4 i2000SR 230 VAC power supply
J5 IM Track CAN-BUS
J6 Future remote display/keyboard
J7 Track CAN bus barcode
J8 LAS i2000SR Interface – (Master tag reader RS-232 port 1) to cable 036810251 to
i2000SR SCC Edgeport Port 4 (Com 6)
J9 Priority, Routine LEDs and Keypad
J10 Electro valves on pneumatic panel
J11 i2000SR IM power switch
J12 Carrier Detector (SICK optical sensor) detects that a carrier was returned to track
and activates the track anti-collision gate (G9)
J13 i2000SR IM G6 G7 G8 D3 Carrier Detector (SICK optical sensor) gates
J14 Optional - To be used to program IM Track barcode reader
J15 Not used. Optional
J16 Not Used
J17 Not Used
J18 i2000SR CLI Interface - (Master tag reader RS-233 port 2) to cable 625810280 to
i2000SR SCC Edgeport port 1 (Com 3)
J19
+24 volt return wire extended from the i2000SR IM via a hole at J19. Routed to
interface with a process path ground wire installed on the i2000SR IM (i2000SR
Ground Wire)

Table of Contents