− Diverter is activated once per carrier.
• Queues are properly built.
− Six (6) carriers at the Priority Gate (G8)
− Eleven (11) carriers at the Routine Gate (G7b)
− Four (4) carriers at Output Gate (G6)
Verify Config Wizard i2000SR Setting
NOTE:
Configuration parameters may change. To inform the field
of the change, a service advisory will be released. To
ensure that the APS is in the latest configuration refer to
the current service documentation release.
Modify Config Wizard for i2000SR
NOTE:
In the following, the symbol # is used to reference the
workcell instrument number of the i2000SR being installed.
• Login as FSE.
• Diagnostics
• Automation
• Blank Button
• Config Wizard
• In the Config Wizard, select the following:
a. Analyzers
Instrument # Type - i2000SR
Configure #
NOTE:
05X is the CAN bus address of the Master Tag Reader in the
IM electronic box. The Tag Reader board has two RS-232
communication ports. :1 is used to indicate that the
i2000SR LAS communication is connected to port 1. Port 2
is not used for LAS communication.
i2000SR CAN # Dip Switch
1 50 2,5,6
2 51 1,2,5,6
3 52 3,5,6
4 53 1,3,5,6