Service Manual i2000SR interface module
625798100.APS.5.doc Page 73 of 212
Diverter is activated once per carrier.
Queues are properly built.
Six (6) carriers at the Priority Gate (G8)
Eleven (11) carriers at the Routine Gate (G7b)
Four (4) carriers at Output Gate (G6)
Verify Config Wizard i2000SR Setting
NOTE:
Configuration parameters may change. To inform the field
of the change, a service advisory will be released. To
ensure that the APS is in the latest configuration refer to
the current service documentation release.
Modify Config Wizard for i2000SR
NOTE:
In the following, the symbol # is used to reference the
workcell instrument number of the i2000SR being installed.
Login as FSE.
• Diagnostics
• Automation
• Blank Button
• Config Wizard
In the Config Wizard, select the following:
a. Analyzers
Instrument # Type - i2000SR
Configure #
NOTE:
05X is the CAN bus address of the Master Tag Reader in the
IM electronic box. The Tag Reader board has two RS-232
communication ports. :1 is used to indicate that the
i2000SR LAS communication is connected to port 1. Port 2
is not used for LAS communication.
i2000SR CAN # Dip Switch
1 50 2,5,6
2 51 1,2,5,6
3 52 3,5,6
4 53 1,3,5,6