625798100.APS.5.doc Page 21 of 212
In the event of a sample presentation error while running in APS
mode, the sample will be flagged with a sample presentation error
and the i2000SR will continue processing. The yellow lamp will
not be illuminated or be blinking.
NOTE:
A Sample Presentation error is generated when the RFID
read at the end of the aspiration differs from that read at
the beginning. If the end RFID reading is missing, a sample
presentation error is generated. If another RFID is read,
the first sample will be flagged with a sample presentation
error and the second carrier will be flagged with a sample
queue error.
When in the Standalone mode, in the event of a sample
presentation error the yellow lamp will begin blinking and the red
lamp will be turned ON. The IM will issue a stop command to the
CLI port to prevent the i2000SR analyzer from processing samples
GREEN LAMP ON: Running, Paused or Pausing
The green lamp is illuminated to indicate that the i2000SR and IM
are operating without error and the communication link between
i2000SR and IM is established.
Modes of Operation
Run:
Standalone mode – the user presses the Run button on the
keypad. Samples loaded at either the Routine or Priority Inputs
with test orders scheduled on the i2000SR analyzer are
processed.
APS mode – the i2000SR instrument(s) is selected to be on-line.
With the appropriate i2000SR IM initialized and analyzer running
(IM pole light Green), the APS will process and route samples with
i2000SR-configured tests to the IM.
Pause:
Standalone mode – the user presses the PAUSE button on the
keypad. The IM track will stop processing samples. Standalone
sample processing will resume when the RUN button on the LUI
panel is pressed.
APS mode – at the APS GUI the user selects Diagnostics, Track,
Transportation Module – RFID Tag Reader Instrument and then
PAUSE. Samples on the IM will stop being processed until the IM
is RESET.
Exercise:
Standalone – at the LUI panel the user presses and holds the RUN
key then presses the Pause key. Off-line carriers are moved within
the IM track. On-line carriers are routed to the main APS track.