EasyManua.ls Logo

Architect i2000SR - Alignment of routine and STAT sampling gates

Architect i2000SR
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
i2000SR interface module Service Manual
Page 176 of 212 625798100.APS.5.doc
Alignment of routine and STAT sampling gates
ALIGNMENT PROCEDURE SCHEDULE no. 16
Operation:
Routine and STAT Sampling gates
alignment
Working area: i2000SR IM Mechanical Group: 625.102.00X
Electrical Group: 625.800.00X Part no. 515.405.00X
Machine Status: Operating in standalone mode
Skill Required: FSE (Field Service Engineer)
Time Required: TBD (To Be Determined)
Special Tools Required: 475.110.031 (oversized carrier tool)
Activity Type: ; Setting Control Cleaning Replacement
Intervention kind: Electrical ; Mechanical Fluidic Pneumatic
Daily Weekly Monthly
Quarterly Half yearly Yearly
Two-year After signalizing Because of failure
Frequency:
; When necessary
WARNING: Potential Biohazard. This is an activity or area
where you may be exposed to potentially infectious
material. Refer to “Section 4 - Hazards and Warnings
” of
the Accelerator Service Manual
Reduce exposure by using proper personal protection
equipment and cleaning surfaces with 0.1% sodium
hypochlorite (bleach).
WARNING: Pinch hazard
WARNING: Probe Sharps Hazard
Follow safe sharp practices. Do not place your hand near
the probe while it is moving. Put the instrument through a
decontamination cycle and wipe the probe with a
disinfectant prior to touching or working near the probe.
GOAL:
During sample aspiration, the sample and STAT probes
must only touch the fluid at the bottom of the tube.
During the z-axis travel the probe must stay centered on
the tube. Adjust the probe(s), gate(s), antenna, analyzer
feet, and support bracket as required to center the probe
at both the upper and lower positions.

Table of Contents