EasyManua.ls Logo

Architect i2000SR - Page 175

Architect i2000SR
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual i2000SR interface module
625798100.APS.5.doc Page 175 of 212
STEPS REFERENCES
10 Verify that the screw shown in the
figure is firmly fastened to avoid
movement of the BCR .
VERIFICATION
STEPS REFERENCES
1 Order tests on five sample tubes for
each barcode type being used.
2 Verify that the barcodes are correctly
read and samples are processed.
3 If i2000SR IM is in standalone mode,
the BCR alignment can be checked
using the Alternate Method procedure
above.
Requires a PC running BARCODE SICK SCANNER
CLV 422-0010 software and a standard serial null
modem cable connected between PC serial port
and J14 on top of the electrical box.
During the procedure, follow software instructions

Table of Contents