Operating the System
Perform examination
4.4.8 Basic exposure error handling
Exposure not possible
Check
Measure
The small detector is se-
lected (Green)
Is the small detector in the
docking station?
Remove the small detector
from the docking station.
Table examination
If the table is equipped with
a wireless detector and
charging the detector in the
holder - check if the connec-
tor is correctly connected to
the detector.
Connect the connector cor-
rectly to the wireless
detector.
Table or wall stand
Is the indication light lit (yel-
low LED in the control room
or green light on the OTC)?
Position the system correct
by pressing the servo button
on the remote control, the
operators console or on the
OTC.
or
Change from Auto to Man-
ual on the OTC display if the
patient/light field is in posi-
tion (and you don´t want to
reposition the OTC).
Table examination Is the detector in the table
detector holder?
Place the detector in the ta-
ble detector holder, make
sure to connect the connec-
tor correctly.
Wall stand examination Is the detector in the wall
stand detector holder?
Place the detector in the
wall stand detector holder,
make sure to connect the
connector correctly.
Operation Manual
1000-095-051-EN Rev. 1.1 ARCOMA AB
2020-06-16 Page 103 SWEDEN