Appendix B
Monthly Checklist
12 Appendix B
12.1 Monthly Checklist
Make a copy of this paper before filling in.
If there is any discrepancy please use the table to make a note.
Hospital:..................................................................................................................
Id no:.................................................... Room:....................................................
Sign:.....................................................
Date:......................................................
12.1.1 OTC
1. Move the OTC manually to all positions in X, Y and Z di-
rection and make sure it runs smoothly and sounds OK.
2. Check the emergency stop. By activating the emergency
stop all motorized movements are inhibited.
3. Choose table position and make sure the tracking is
activated.
Measure between the X-ray tube focal spot and the ac-
tive image receptor surface of the image receptor holder.
The measured SID shall correspond with the displayed
SID.
Move the OTC in X or Y direction and measure between
the X-ray tube focal spot and the active detector surface
of the detector holder again. The SID is allowed to differ
±1%.
4. Check that the SID, shown on the display of both the Im-
age system and the collimator, correspond with the
measured SID.
5. Check the hoses for damage.
6. Check all outer cabling for damage.
7. Clean all outer surfaces, except for the lubricated column
segments. See Chapter “Cleaning” at the Instruction for
use