Function and safety checks
Monthly Checks
7 Function and safety checks
7.1 Monthly Checks
7.1.1 General
If any malfunction is detected, the entire equipment must be taken out of use until the prob-
lem is eliminated by a service engineer from the supplier or by the local technical staff trained
by the supplier.
Note!
Before performing any maintenance please read the Safety chapter.
Daily and monthly checks are normally performed by the user/operator.
Annual checks shall be performed either by local technical staff trained by the supplier or au-
thorized service representatives.
Checks for all units.
OTC, table and Wallstand:
1. Check the cable hose for damage.
2. Check all outer cabling for damage.
3. Clean all outer surfaces, except for the lubricated column
segments.
4. Make sure that the Instruction for use is available and up to
date.
5. Check the emergency stop. By activating the emergency stop
all motorized movements are inhibited. See Safety Chapter
for information on how the Emergency stop should react on
command.
7.1.2 OTC
6. Power up the OTC and check all functions.
7. Move the OTC around and observe any irregularities.
7.1.3 Table
8. Move the Table in X, Y and Z direction to make sure it runs
smoothly and sounds OK.
9. Move the table top longitudinal and check that the mechanical
end stops are not loose.
7.1.4 Wallstand
10.Move the Wallstand up and down in Z direction and make
sure it runs smoothly and sounds OK.
Operation Manual
1000-095-051-EN Rev. 1.1 ARCOMA AB
2020-06-16 Page 145 SWEDEN