Safety
Service and Maintenance
2.4 Service and Maintenance
WARNING!
When service or maintenance will be performed the technician shall lock the equip-
ment from all energy sources.
Be aware of that there will be live parts even some time after having switched off the
mains.
Due to remaining energy, always wait at least 5 minutes before working on the
System.
WARNING!
There will still be live parts even when the System is switched off.
WARNING!
The equipment must not be serviced or maintained while in use with the patient.
Risk for personal injury.
Service and maintenance shall only be performed when patient is not present.
The equipment must be checked according to the 7 Function and safety checks to maintain
reliability and serviceability, and to ensure the safety of the patients, the operator and third
parties.
If national rules or regulations specify more frequent checks and/or maintenance, such regu-
lations must be observed.
2.4.1 Operation, Installation and Repair
WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
WARNING!
Only medical-approved products shall be in the X-ray room
Risk of electric shock to patient or user
No non-medical electrical devices shall be used in the x-ray room.
Note that the monitor and the PC for the Image system, are none-medical approved
products.
ARCOMA AB 1000-095-051-EN Rev. 1.1
Operation Manual
SWEDEN Page 14 2020-06-16