Operating the System
Movement Short-cut Zones
4.5 Movement Short-cut Zones
A short-cut movement is defined as, when the OTC moves directly to the intended position,
without first moving up into the transport interval zone.
Movement short-cut zones are defined both for the Table and the Wallstand, as rectangular
cubes. (Short-cut zones can be defined by a servcie engineer)
To perform a short-cut movement, the following conditions must be fulfilled;
The System shall be positioned above the Table, or at the specified short-cut zone in front
of the Wallstand.
At least one of the OTC corners must be inside the short-cut zone at the start of the
movement.
Outside these short-cut zones, the OTC will first move in Z direction, to the transport inter-
val zone, before moving to position.
ARCOMA AB 1000-095-051-EN Rev. 1.1
Operation Manual
SWEDEN Page 104 2020-06-16