User Interface
Automatic Collimator Control
Service Display
SETTINGS
LOG
DISPLAY
SERVICEUSER SETTINGS
Versions
KERNEL
2.6.37-14321-g1fb710c
ROOTFS
ME Merisc (Poky 8.0 base)
U-BOOT
2010.12-rc2-00004-g71lede39
MLO
X-Loader 1.44 (ME)
Protocol
01.01
System
1.123.1234.1245
GUI
1.1 (Oct 7 2013 08:56:26)
3593
Fig. 3-21 Service Display
Information of the display software versions.
3.2.7 System Mode
The System has a number of different modes. All modes are described below with their spe-
cial functionalities and features.
Note that depending on the particular System, different modes and actual configurations are
available.
Free mode
Auto position mode
Wall flexible mode
Table flexible mode
Film tracking mode
Pendulum mode, Table
Stitching mode (toward the Table and Wall stand)
3.2.7.1 Free Mode
General Description
The Free mode is the most basic mode in the System.
The mode holds no special features or functionality. It is intended as a manual mode with a
lot of freedom in positioning and exposure, e.g. for emergency examinations or examinations
with the patient in a wheel chair or lying in a bed. This mode is available in all Systems.
The distance shown in display for Free mode is the distance to the floor.
Exposure Validation
Exposure is allowed if the stand is not moving and operating properly (not in an error state).
Operation Manual
1000-095-051-EN Rev. 1.1 ARCOMA AB
2020-06-16 Page 55 SWEDEN