User Interface
Automatic Collimator Control
3.2.7.2 Auto Positioning Mode
General Description
The Auto position mode functions as the Free mode with the added functionality of automatic
positioning in the room.
Automatic positions are a number of positions that can be programmed and saved into the
System. The stand will upon activation of the servo button, move to the programmed position
chosen from the imaging unit.
The mode is intended as a flexible, easy to use mode. The mode can also be used as a park-
ing mode.
The distance (H) shown in display for Auto position mode is the distance to the floor.
Exposure Validation
Exposure is allowed if the stand is not moving and operating properly (not in an error state).
The chosen position must have been reached successfully.
ARCOMA AB 1000-095-051-EN Rev. 1.1
Operation Manual
SWEDEN Page 56 2020-06-16