Operating the System
Manual Mode
4.7.1.1 Deactivation of Manual Mode
The Manual mode is deactivated by selecting a new APR with a different auto-position. De-
activation is also done by changing to Automatic mode .
If the same patient is examined and an APR with the same auto-position is selected, the
Manual mode will be kept.
4.7.1.2 Restrictions in Manual Mode
In Manual mode the shown SID value is based on the assumption that the detector is posi-
tioned in the same way as it should have been in an active servo mode. Note that the SID
value will not be able to be calculated is some positions. No SID value will be shown if the
System is not in an active servo mode when Manual mode is activated.
Tracking or other mode movements are not possible when Manual mode is active.
ARCOMA AB 1000-095-051-EN Rev. 1.1
Operation Manual
SWEDEN Page 108 2020-06-16