Safety
Electromagnetic compatibility (EMC)
Guidance and manufacturers declaration - electromagnetic immunity
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Note!
U
T
is the AC mains voltage prior to application of the test level.
Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than
the recommended separation distance,
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance;
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
6 Vrms (ISM and
amateur radio
bands)
3 Vrms
150 kHz to 80
MHz
6 Vrms (ISM and
amateur radio
bands)
d = 1.2 √p
Radiated RF IEC
61000-4-3
3 V/m
10 V/m
80 MHz to 2.7
GHz
3 V/m
10 V/m
80 MHz to 2.7
GHz
d = 1.2 √p 80 MHz to 800 MHz
d = 2.3 √p 800 MHz to 2.7 GHz
where p is the maximum output power rat-
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d is
the recommended separation distance in
metres (m).
Proximity field
from wireless
transmitters
61000-4-3
9 V/m to 28 V/m
15 specific
frequencies
9 V/m to 28 V/m
15 specific
frequencies
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
NOTE 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
ARCOMA AB 1000-095-051-EN Rev. 1.1
Operation Manual
SWEDEN Page 36 2020-06-16