Notice that <Area1> is the sensor name of an Area Sensor Type in the current inspection.
>> get blemish_history <Blemish1> minedgelength\x0D\x0A
<< OK\x0D\x0A
<< 22\x0D\x0A
Notice that <Blemish1> is the sensor name of a Blemish Sensor Type in the current inspection.
8.4.6 Command Channel Command Status Register
The command status is a verification of the command type.
Command Status Value (16-bit integer)
Unknown 0
Read 1
Write 2
Execute 3
8.4.7 Command Channel Error Codes
Table 13: Plus TG Command Channel Error Codes
Numeric ID Text ID Description
00000 SUCCESS Command processed successfully
10000 EMPTY_FRAME_RECEIVED Indicates that the request was empty. The command channel
requires a command, any arguments, and an end-of-frame
delimiter.
10001 COMMAND_NOT_RECOGNIZED The command specified is not recognized
10100 GROUP_MISSING A Group ID must be specified immediately after the command
10101 GROUP_NOT_FOUND The specified Group ID is invalid / unknown
10102 GROUP_ITEM_MISSING A Group Item ID must be specified immediately after the Group
ID
10103 GROUP_ITEM_NOT_FOUND The specified Group Item ID is invalid / unknown
10152 NOT_READABLE Attempt to get a value that is not readable
10153 NOT_WRITEABLE Attempt to set a value that is not writeable
10250 NOT_A_METHOD Method ID specified is not a method
10251 WRONG_ARGUMENT_COUNT Total method arguments specified do not match method
10252 COMMAND_NOT_FINISHED Attempt to issue command when a previous command has not
finished
10300 INVALID_ARGUMENT_TYPE Item ID specified must be a item (not a group or method)
10301 DATA_VALUE_MISSING Command missing item's data value
10350 ARGUMENTS_DETECTED Get command received with unneeded arguments
10351 INVALID_ARGUMENT_TYPE Item ID specified must be a item (not a group or method)
10340 MINIMUM_VALUE_EXCEEDED New item value is below the minimum
10341 MAXIMUM_VALUE_EXCEEDED New items value is above the maximum
10500 DATA_SET_EMPTY Data export operation returned no results.
10900 SENSOR_NOT_READY Command specified requires sensor to be in the READY state.
10920 SENSOR_TYPE_NOT_ACTIVE Command specified belongs to a different sensor type.
15000 VALUE_INVALID Text value is invalid / unknown
15050 VALUE_INVALID Text value is invalid - expecting True or False
15100 STRING_TOO_LONG String value specified exceeds maximum allowable length
iVu Plus TG Gen2 Image Sensor
98 www.bannerengineering.com - Tel: 763.544.3164