SYSTEM CONFIGURATION Chapter 8
8 - 1
SYSTEM CONFIGURATION
8.1 GENERAL
System configuration modes include Configuration Review mode and Configuration Set
mode.
NOTE
For a complete listing of the factory default configurations, see the
tables at the end of this chapter.
8.2 CONFIGURATION REVIEW MODE
Configuration Review mode allows review of configurable options without risk of altering
them.
8.2.1 Activating Configuration Review Mode
To activate Configuration Review mode, press the key sequence 1,2,3 following completion
of the POST.
8.2.2 Configuration Review Sequence
Following the software version display, the pump prompts for each group of review options
by displaying VIEW <group name>. Y is displayed in the syringe manufacturer field. To
access the indicated option group, press CONFIRM. To skip to the next group press ▲ or
▼ to change the Y to N and press CONFIRM. Continue to press CONFIRM to view each
option in a group until the next group prompt appears. Again, use the ▲ or ▼ key to view
that group’s options.
When the last configuration group option has been reviewed, the pump automatically
restarts the POST. Also, switching the pump “OFF” any time during the review will cancel
the review process.
8.3 CONFIGURATION SET MODE
Configuration Set mode is primarily used to select and control configurable options, and
provides the following functions:
• create, modify and delete protocol libraries, and control availability of specific protocol
libraries
• create, modify and delete protocols
• control access to LIBRARY group functions
• transfer (replication) configuration from pump to pump
• control availability of specific infusion modes
• control availability of specific syringe manufacturers
• control availability of specific syringe sizes