Chapter 2 THEORY OF OPERATION
2 - 26
Complete remote monitoring and control application information is provided in the Auto
Syringe® AS50 Infusion Pump Programmer’s Manual.
2.3.4.2 Event Recording Feature
The pump has the capability of recording 253 events in EEPROM. The pump maintains
two separate logs: the event log and the fail-safe log. The event log has a capacity of 243
events and the fail-safe log, 10 events. Each log is maintained as a circular list; that is,
when the log is full, the oldest event is replaced with the next event. The logs are defined
and accessed by way of the serial port using instructions outlined in the Auto Syringe®
AS50 Infusion Pump Programmer’s Manual.
2.3.4.2.1 Dumping the Event Log
The Serial Communication Feature will provide remote commands for transmitting the
contents of the fail-safe alarm log and the event log to a host computer. Refer to the
Auto
Syringe®
AS50
Infusion
Pump
Programmer’s
Manual
for additional information.
2.3.5 Displaying Additional Program Data
The following sequence will be followed when CONFIRM is pressed either during Run or
Standby state to display additional data:
1. The battery voltage will be displayed in the text field for 3 seconds.
2. Next, the operating mode will be displayed in the text field for 3 seconds.
3. The normal display will be restored.
NOTES
If the pump is in Run state, the SYRINGE mfr/size field will be
displayed during steps 1 - 3.
If the pump is in CUSTOM DILUTION mode, the original drug
concentration, FINAL VOL and DRUG VOL fields will be displayed
during steps 1 - 3.
Pressing any key during the above sequence will abort the sequence
and restore the normal display.
2.3.5.1 Battery Voltage Display
The battery voltage will be displayed when CONFIRM is pressed for displaying additional
program data. Line 1 of the text field will display the voltage (for example “7.0 V”). Line 2
will display BATTERY.
If the CONFIRM key is pressed during the lamp test portion of the POST, the battery
voltage will be displayed for two seconds immediately following the lamp test.
2.3.6 Specific Errors, Codes, and Alarms

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals