FUNCTIONAL TESTS Chapter 4
4 - 5
4.2.8 Keypad Check: (*D*)
1. (*D*) Press and briefly hold down the START and STOP keys one at a time. The
pump should sound repeated beeps while each key is held down.
NOTE
Holding down the START key for a prolonged period of time may
cause a “stuck key” fail-safe alarm requiring a system restart.
2. (*D*) Press every key, except the arrow keys, in any order, one key at a time. Each
key press should result in a short beep.
3. (*D*) Press any two or more keys simultaneously. There should be no response.
4. (*D*) Press and hold any key except the arrow keys, the START key, or the STOP
key. There should be only one beep, no matter how long the key is held down.
5. (*D*) Press and hold any key except arrow keys, START, or STOP keys. Then press
any other key. There should be no response to the second key press.
4.2.9 Range of Motion Check: (*D*)
1. (*D*) Check the plunger driver and the barrel clamp for free travel through their full
range.
2. (*D*) Check that the plunger seats properly and that the plunger driver does not slip on
the leadscrew when the plunger is fully closed.
3. (*D*) Check that both barrel clamp release levers operate properly.
4. (*D*) Check the tab lock to be sure it opens when the barrel clamp is fully opened and
closes when the barrel clamp is fully closed.