PRODUCT UPDATES APPENDIX D
D - 1
D.1 GENERAL
This appendix describes major updates to the pump. The updates are listed by serial
number and/or software revision number of when the change occurred. In many instances
a different part number is assigned for compatibility reasons. To ensure proper fit and
operation of parts, make sure that you check all updates that may apply to a particular
serial number.
CHANGES TO SERVICE MANUAL ARTWORK REV. A INCLUDE THE FOLLOWING:
1. Change in software version from V7.00.04 to V7.01.05. This includes changes to the
Master Processor error codes listed in section 6.4. It also allows the date and time to
be replicated from a source pump.
2. The Factory Default Configuration changed as follows (refer to section 8.6):
2.1 The “LIBRARY” option in the MISC Group changed from Y to N.
2.2 The “ENABLE DB9 PORT” option in the MISC Group changed from N to Y.
2.3 The IDENT Group identifier changed from “THANKYOU” to “AS50.”
3. The acceptable occlusion range changed from 13.5 – 17.5 psi to 12.0 – 20.0 psi (refer
to section 4.2).
4. The Molded End Lever design was modified (refer to sections 7.2 and C.2). The
original lever was mounted with a 4-40 x 1/4" flat head screw. The revised lever is
mounted with a 4-40 x 5/16" flat head screw and a hex nut.
5. The recommended battery charge time was changed from 24 hrs. to 16 hrs. (refer to
sections 4.1 and 6.2).