ALERTS AND ALARMS Chapter 10
10 - 5
If the alarm occurs before any data is received, it may be cleared by pressing any key. The
pump returns to the SELECT FUNCTION prompt, and the RECEIVE function may be
selected again.
If the alarm occurs while configuration replication is in progress, a keypress will silence the
audio portion of the alarm only. The text message and ALERT LED persist, and the only
recovery is to cycle power. Further, the destination pump will display CONFIG CORRUPT
following the next power up. A valid configuration must be loaded into the pump before it
may be operated normally.
10.3.4 Comm Timeout
Cause: interrupted data flow during configuration replication
Audio response: continuous beeps
LED response: ALERT LED flashes while message is displayed
Message duration: until <5089> code is entered
Issued during configuration replication if the destination pump detects an interruption in
data flow for more than 5 seconds. The pump drops out of replication.
If the alarm occurs before any data is received, it may be cleared by pressing any key. The
pump returns to the SELECT FUNCTION prompt, and the RECEIVE function may be
selected again.
If the alarm occurs while configuration replication is in progress, a keypress will silence the
audio portion of the alarm only. The text message and ALERT LED persist, and the only
recovery is to cycle power. Further, the destination pump will display CONFIG CORRUPT
following the next power up. A valid configuration must be loaded into the pump before it
may be operated normally.
10.3.5 Config Corrupt
Cause: library or configuration data corrupted
Audio response: single beep
LED response: none
Message duration: until <9805> is entered or power is cycled
Issued following the power-on self test (POST) if the pump detects during the POST that
library or configuration data is corrupted. Pressing the key sequence 9,8,0,5 is the only
valid operator action.
Also issued if the pump detects corrupted data during protocol selection. In this instance,
the pump will not respond to any further operator action, and will issue a CONFIG
CORRUPT error following the POST on the next power up.