SYSTEM CONFIGURATION Chapter 8
8 - 11
Refer to Chapter 9, CONFIGURATION REPLICATION.
8.4.2 MODES Group
The MODES configuration group allows the operator to individually enable and disable
each of the 18 infusion modes. Only enabled modes are displayed at the SELECT
MODES programming step. To select the MODES group confirm a Y response to the
CONFIG MODES prompt.
Each of the eighteen infusion modes is indicated by its normal Standby state Mode
Selection display. Each mode display includes a Y or N response indicating respectively
that the mode is currently enabled or disabled. For each infusion mode, press ▲ or ▼ if
necessary to change the response and press CONFIRM to complete the setting. At least
one mode must be enabled. When CONFIRM is pressed for the last mode display, the
pump displays UPDATING.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
8.4.3 SYRINGES Group
The SYRINGES configuration group allows the operator to individually enable and disable
each of the three syringe manufacturers. To select the SYRINGES group confirm a Y
response to the CONFIG SYRINGES prompt.
The SYRINGES group presents a series of syringe manufacturer displays:
1. ENABLE B-D = Becton Dickinson®
2. ENABLE MONOJECT = Monoject®
3. ENABLE TERUMO = Terumo®
Each syringe manufacturer display includes a Y or N response, indicating respectively that
the manufacturer is currently enabled or disabled. Only enabled manufacturers are
displayed at the SELECT MFR programming step. For each manufacturer, press ▲ or ▼
if necessary to change the response and press CONFIRM to complete the setting. At least
one syringe manufacturer must be enabled. If a N response is confirmed at a syringe
manufacturer display, the pump proceeds to the next manufacturer display.
If a Y response is confirmed at a syringe manufacturer display, the pump displays a
SELECT SYRINGES prompt with a response of Y. Press CONFIRM to enable or disable
individual syringe selections, or press ▲ or ▼ to change the response to N and press
CONFIRM to continue with the next manufacturer display.
If a Y response is confirmed at the SELECT SYRINGES prompt, the pump displays a
series of syringe size selections. Each selection is presented with a Y or N response,