Chapter 8 SYSTEM CONFIGURATION
8 - 2
• enable/disable “Default to Previous” feature for protocol library, protocol, mode, and
syringe manufacturer
• enable/disable Remote Control feature
• enable/disable Protocol Library feature
• enable/disable Size Override feature
• control occlusion detection sensitivity
• control audio volume range
• control maximum rate of infusion
• enable/disable Auto Lock feature
• enable/control/disable backlight
• select serial port
• enable/disable PUMP IS IDLE alarm feature
• enable/control/disable Near End alert
• enable/disable BOLUS COMPLETE alert
• enable/disable DOSE COMPLETE alarm
• enable/disable Volume Limit feature for ML/HR and ML/MIN modes
• define alphanumeric identifier
• set date/time feature
8.3.1 Activating Configuration Set Mode
To activate Configuration Set mode, press the key sequence 9,8,0,5 following completion
of the POST.
NOTE
The pump sounds an error beep for each key. These tones are
provided to discourage unauthorized access and may be ignored.
8.3.2 Date and Time Display
On entry to Configuration Set mode the pump initially displays the date and time in the
following format:
12.00AM
12/13/95
8.3.3 Software Version Display
Following the date and time display, the pump presents the software version display.
The first line of the display indicates the pump software version. The second line is divided
into two fields. The left field displays the configuration version, a 2-character code that
identifies the format of the configuration data. The right field displays the 4-character
configuration checksum, which identifies the entire current configuration file including the
protocol library.