ALERTS AND ALARMS Chapter 10
10 - 1
10.1 GENERAL
This chapter contains detailed descriptions of alert and alarm conditions not listed in the
pump Operator’s Manual.
10.2 ALERTS
10.2.1 Cal Invalid
Cause: invalid calibration reading
Audio response: single beep
LED response: none
Message duration: 2 seconds
1. Issued during any calibration if the second reading taken is not smaller than the first, or
if either calibration measurement is grossly out of range. Following the alert the pump
restarts the calibration sequence.
2. Issued at the completion of pressure sensor calibration if the force equivalent of the
maximum A/D reading (1023) is not at least 18 pounds. Following the alert the pump
restarts the calibration sequence.
10.2.2 Code Mismatch
Cause: invalid library access code entered
Audio response: single beep
LED response: none
Message duration: 2 seconds
1. Issued if an invalid library access code is entered following the ENTER CODE prompt
at the start of protocol library editing. Following the alert the pump redisplays ENTER
CODE.
2. Issued during the CHANGE CODE function when the new LIBRARY group access code
is entered for the second time and the second code entry does not match the first.
Following the alert the pump redisplays the <code> CODE prompt.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals