Chapter 2 THEORY OF OPERATION
2 - 16
2.3 SOFTWARE
2.3.1 Sensor Calibration
Sensor calibration modes include Calibration Review mode and Calibration Set mode. See
Chapter 5 for complete instructions regarding sensor calibration.
2.3.1.1 Calibration Review Mode
Calibration Review mode allows review of the calibration settings without the risk of altering
them.
Calibration Review mode allows the operator to review the following sensor calibration
settings (in the following sequence):
1. Plunger Driver Position Sensor
2. Syringe Recognition Sensor
3. Pressure Sensor
4. VB_SENSE
In addition, Calibration Review mode provides an auxiliary LCD test for examining each
LCD segment individually.
2.3.1.1.1 Activating Calibration Review Mode
To activate Calibration Review mode press the key sequence 3,2,1 following completion of
the POST (Power-on Self Test).
2.3.1.1.2 Calibration Review Sequence
1. Plunger Driver Position Calibration Review
During plunger driver position calibration review, the pump continuously displays the
analog and digital plunger driver position sensor readings, based on the current calibration
settings. The text field displays DRIVEPOS.
Press CONFIRM to continue. The pump clears the readings, and continues with the
syringe recognition calibration review.
2. Syringe Recognition Calibration Review
During syringe recognition calibration review, the pump continuously displays the analog
and digital barrel clamp position sensor readings, based on the current calibration settings.
The text field displays SYRNGREC.
Press CONFIRM to continue. The pump clears the readings, and continues with the
pressure sensor calibration review.