Chapter 6 TROUBLESHOOTING AND REPAIR
6 - 2
6.2.2 Troubleshooting Tips
TABLE 6-1
Symptom Possible Cause Solution
Constant CHECK SYRINGE
alarm (This message
indicates two or more
mounting points (flange,
barrel, plunger) are not
correctly positioned.)
Improperly loaded syringe Reload syringe.
Pump out of calibration Check calibration and re-
calibrate, if necessary.
Contaminated barrel clamp Perform service cleaning
(see Chapter 5).
Malfunctioning barrel clamp
and/or barrel clamp zebra
strip
Replace barrel clamp and/or
barrel clamp zebra strip and
recalibrate.
Worn, frayed or damaged
barrel clamp zebra strip
Replace barrel clamp zebra
strip and recalibrate.
Syringe recognition will not
calibrate
Contaminated barrel clamp Perform service cleaning
(see Chapter 5).
Malfunctioning barrel clamp
and/or barrel clamp zebra
strip
Replace barrel clamp and/or
barrel clamp zebra strip and
recalibrate.
Worn, frayed or damaged
barrel clamp zebra strip
Replace barrel clamp zebra
strip and recalibrate.
Intermittent EMPTY alarm
Pump out of calibration Check calibration and re-
calibrate, if necessary.
Intermittent BATTERY LOW
alarm
Battery may require charging Recharge unit for a
minimum of 16 hours and
verify battery operation.
Loose battery connection Open battery compartment
and check battery
connection.
Loose charger connection Wiggle the charger
connector while plugged

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals