Chapter 8 SYSTEM CONFIGURATION
8 - 16
press. Press CONFIRM to accept the current setting or enter a new value of 0 to 60
minutes and press CONFIRM to complete the setting.
NOTES
If the pump goes into a Low Battery condition, the backlight ‘on’ time is
temporarily reduced to a maximum of 2 minutes for each key press.
If 0 minutes is configured, the backlight is not activated unless the
charger is connected.
If the charger is connected, the backlight is kept on during the entire
POST, but is turned off for 1/2 second at the start of the first
programming prompt.
8.4.5.9 ENABLE DB9 PORT
The ENABLE DB9 PORT option determines whether the DB9 port or the mini-DIN port will
be used as the serial I/O port. This option allows one of two possible responses: Y or N.
Y The DB9 port will be used as the serial I/O port.
N The mini-DIN port will be used as the serial I/O port.
NOTE
Replication cables for use with the DB9 port (catalog no. 2M8155) are
available. See Chapter 9 for the replication procedure.
8.4.6 ALERTS Group
The ALERTS group includes the following options:
• IDLE ALARM
• SET NEAR END
• BOLUS COMPLETE
• DOSE COMPLETE
• VOLUME LIMIT
To select the ALERTS group confirm a Y response to the CONFIG ALERTS prompt.
Unless otherwise instructed, for each option or prompt, press ▲ or ▼ if necessary to
change the response and press CONFIRM to complete the setting.