SYSTEM CONFIGURATION Chapter 8
8 - 15
Syringe Configuration Setting
Size L M H
60 8 13 17
30,35 8 14 19
20 9 15 20
10,12 10 15 21
5,6 12 17 24
3 14 20 27
1 19 27 27
8.4.5.5 AUDIO RANGE
The AUDIO RANGE option determines the audio circuit output level. This option allows
one of two possible responses: L (low) or H (high).
8.4.5.6 RATE RANGE
The RATE RANGE option determines the maximum infusion rate. This option allows one
of three possible responses:
L (low) maximum rate = 15mL /hr (0.25 mL/min)
M (medium) maximum rate = 120 mL/hr (2 mL/min)
H (high) maximum rate = approximately 366 mL/hr or 6.1 mL/min for a B-D® 60 mL
Plastic syringe
8.4.5.7 AUTO LOCK
The AUTO LOCK option determines whether the Auto Lock feature is enabled. This option
allows one of two possible responses: Y or N.
Y The Auto Lock feature is enabled.
N The Auto Lock feature is disabled.
8.4.5.8 SET BACKLITE
The SET BACKLITE option allows the operator to determine the operation of the backlight
when pump power is on and the charger is not connected.
The pump displays SET BACKLITE with a N response. Press CONFIRM to skip to the
next option or press ▲ or ▼ to
change the response to Y and press CONFIRM to access
the backlight setting.
If Y is selected, the pump displays a <xx> MINUTES prompt where <xx> is the current
setting. When the charger is not connected, the pump turns on the backlight for the
configured time, following any key press. The timer is restarted at each subsequent key