SYSTEM CONFIGURATION Chapter 8
8 - 13
8.4.4.2 DFAULT PROTOCOL
The DFAULT PROTOCOL option determines the response of the pump to the first arrow
key press following the ▲ ▼ SELECT PROTOCOL programming prompt.
Y The pump displays the name of the last used protocol selection if the last used library is
selected, else it displays the first protocol selection (in the alphabetical sequence) in the
selected library.
N The pump displays NONE if the NONE protocol selection option is enabled, else it
displays the first protocol selection (in the alphabetical sequence) in the selected library.
8.4.4.3 DFAULT MODE
The DFAULT MODE option determines the response of the pump to the first arrow key
press following the SELECT MODE ▲ ▼ programming prompt.
Y The pump presents the mode display for the last used mode if no protocol has been
selected or a protocol with ALL MODES has been selected.
N The pump presents the mode display for the first enabled mode selection if no
protocol has been selected or a protocol with ALL MODES has been selected.
8.4.4.4 DFAULT MFR
The DFAULT MFR option determines the response of the pump to the first arrow key press
following the SELECT MFR ▲ ▼ programming prompt.
Y The pump presents the syringe manufacturer display for the last used syringe
manufacturer if it is enabled, else it presents the syringe manufacturer display for the
first enabled syringe manufacturer.
N The pump presents the syringe manufacturer display for the first enabled syringe
manufacturer.
8.4.5 MISC Group
The MISC (miscellaneous) group displays the following options:
• REMOTE CONTROL
• LIBRARY
• SIZE OVERRIDE
• PSI RANGE
• AUDIO RANGE
• RATE RANGE
• AUTO LOCK
• BACKLITE