Chapter 8 SYSTEM CONFIGURATION
8 - 14
• ENABLE DB9 PORT
To select the MISC group, confirm a Y response to the CONFIG MISC prompt. Unless
otherwise instructed, for each option or prompt press ▲ or ▼ if necessary to change the
response and press CONFIRM to complete the setting.
8.4.5.1 REMOTE CONTROL
The REMOTE CONTROL option determines whether remote control operation is allowed.
This option allows one of two possible responses: Y or N.
Y Remote control operation may be enabled and activated.
N Remote control operation is not allowed.
8.4.5.2 LIBRARY
The LIBRARY option determines whether the Protocol Library feature is enabled. This
option allows one of two possible responses: Y or N.
Y The Protocol Library feature is enabled.
N The Protocol Library feature is disabled.
8.4.5.3 SIZE OVERRIDE
The SIZE OVERRIDE option allows the operator to:
• Override syringe recognition by using the arrow keys to select a syringe size other than
the size displayed by the pump.
• Override a SIZE MISMATCH alert by using CONFIRM.
This option allows one of two possible responses: Y or N.
Y The SIZE OVERRIDE feature is enabled.
N The SIZE OVERRIDE feature is disabled.
8.4.5.4 PSI RANGE
The PSI RANGE option determines the occlusion pressure threshold. This option allows
one of three possible responses: L (low), M (medium) or H (high), which enable specific
fixed alarm threshold points.
Anticipated nominal occlusion pressures in PSI are: