Chapter 9 CONFIGURATION REPLICATION
9 - 4
9.2.4 Error Recovery
If the destination pump alarms because replication was interrupted, perform the following
procedure.
1. Turn off both pumps.
2. Turn on the destination pump. Following the POST, the pump will issue a CONFIG
CORRUPT alarm.
3. Press ā9805.ā The pump will present a modified Software Version display. The
modified Software Version display will indicate the Software Version and the first
character of the Configuration Version, followed by a dash (for example, [A-]). The
Configuration Checksum will not be displayed.
4. At the ENTER CODE prompt, enter the library access code (the default factory access
code is 0000). Press CONFIRM.
5. The destination pump will automatically select the RECEIVE function and display
PRESS START TO RECEIVE OR CLR TO QUIT. DO NOT press any key until told to
do so.
6. Continue at step 1 of the REPLICATION procedure.