Chapter 2 THEORY OF OPERATION
2 - 20
2. The BATTERY LED is on when the plunger capture detector is detecting a syringe
plunger.
3. The STANDBY LED is on when the flange sensor is detecting a syringe flange.
4. The LOCK LED is on when the barrel sensor is detecting a syringe barrel.
2.3.2 Configuration Replication
The Configuration Replication feature allows the entire configuration file including protocol
libraries and the library access code to be copied from one pump (the source pump)
directly to a second pump (the destination pump). This permits replication of configuration
information from a master pump to multiple pumps.
2.3.3 System Configuration
System configuration modes include Configuration Review mode and Configuration Set
mode. See Chapter 8 for complete instructions regarding system configuration.
2.3.3.1 Configuration Review Mode
Configuration Review mode allows review of configurable options without risk of altering
them.
2.3.3.1.1 Activating Configuration Review Mode
To activate Configuration Review mode, press the key sequence 1,2,3 following completion
of the POST.
2.3.3.1.2 Configuration Review Sequence
Following the software version display, the pump prompts for each group of review options
by displaying VIEW <group name>. Y is displayed in the syringe manufacturer field. To
access the indicated option group, press CONFIRM. To skip to the next group press ▲ or
▼ to change the Y to N and press CONFIRM. Continue to press CONFIRM to view each
option in a group until the next group prompt appears. Again, use the ▲ or ▼ key to view
that group’s options.
When the last configuration group option has been reviewed, the pump automatically
restarts the POST. Also, switching the pump “OFF” any time during the review will cancel
the review process.
2.3.3.2 Configuration Set Mode
Configuration Set mode is primarily used to select and control configurable options, and
provides the following functions: