THEORY OF OPERATION Chapter 2
2 - 21
• create, modify and delete protocol libraries, and control availability of specific protocol
libraries
• create, modify and delete protocols
• control access to LIBRARY group functions
• transfer (replication) configuration from pump to pump
• control availability of specific infusion modes
• control availability of specific syringe manufacturers
• control availability of specific syringe sizes
• enable/disable "Default to Previous" feature for protocol library, protocol, mode, and
syringe manufacturer
• enable/disable Remote Control feature
• enable/disable Protocol Library feature
• enable/disable Size Override feature
• control occlusion detection sensitivity
• control audio volume range
• control maximum rate of infusion
• enable/disable Auto Lock feature
• enable/control/disable backlight
• select serial port
• enable/disable PUMP IS IDLE alarm feature
• enable/control/disable Near End alert
• enable/disable BOLUS COMPLETE alert
• enable/disable DOSE COMPLETE alarm
• enable/disable Volume Limit feature for ML/HR and ML/MIN modes
• define alphanumeric identifier
• set date/time feature
2.3.3.2.1 Activating Configuration Set Mode
To activate Configuration Set mode press the key sequence 9,8,0,5 following completion of
the POST.
During the Configuration Set mode, the screen will display [UPDATING] at various times to
store the currently requested changes.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump or press any key while UPDATING is being displayed.