Chapter 8 SYSTEM CONFIGURATION
8 - 18
8.4.6.5 VOLUME LIMIT
The VOLUME LIMIT option determines whether the VOLUME LIMIT alarm is enabled.
This option allows one of two possible selections: Y or N.
Y The VOLUME LIMIT alarm is enabled.
N The VOLUME LIMIT alarm is disabled.
8.4.7 CLOCK Group
The CLOCK group allows the operator to set the real-time clock.
To select the CLOCK group confirm a Y response to the CONFIG CLOCK prompt.
The pump displays the time and date in the text field in the following format (typical date
and time):
12.00AM
12/13/95
To set the time and/or date, press EDIT as necessary to select the desired field. The time
and date fields are accessed in the following sequence: hours, minutes, AM/PM, month,
day, year.
Enter new values for hours, minutes, month, day, or year directly from the keypad. Press
▲ or ▼ to toggle the AM/PM field.
Press CONFIRM to complete entry of the combined date and time.
8.4.8 IDENT Group
The IDENT group allows the operator to select an alphanumeric identifier of up to eight
characters to be displayed following the Power-on self test (POST).
To select the IDENT group, confirm a Y response to the CONFIG IDENT prompt.
Press CONFIRM to accept the current identifier, or change the identifier as indicated in
section 8.5, Entering Text Data. Press CONFIRM to complete the entry.