ALERTS AND ALARMS Chapter 10
10 - 3
10.2.7 Name Invalid
Cause: “NONE” or all blanks entered for library or protocol name
Audio response: single beep
LED response: none
Message duration: 2 seconds
Issued during the MODIFY PROTOCOL, CREATE PROTOCOL, MODIFY LIBRARY or
CREATE LIBRARY function if all blanks or “NONE” is entered as the protocol name.
Following the alert the pump restores the text field and editing cursor.
10.2.8 No Library
Cause: selected function invalid because no library exists
Audio response: single beep
LED response: none
Message duration: 2 seconds
Issued if the MODIFY PROTOCOL, DELETE PROTOCOL, MODIFY LIBRARY, DELETE
LIBRARY, CREATE PROTOCOL or VIEW LIBRARY function is selected and no library
has been created.
Following the alert the pump returns to the ▲▼ SELECT FUNCTION prompt.
10.2.9 Not Empty
Cause: attempt to delete non-empty library
Audio response: single beep
LED response: none
Message duration: 2 seconds
Issued during the DELETE LIBRARY function if the selected (or only) library is not empty.
Following the alert the pump returns to the ▲▼ SELECT LIBRARY prompt.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals