Chapter 8 SYSTEM CONFIGURATION
8 - 10
The CREATE LIBRARY function allows the operator to create a new, empty protocol
library.
The pump displays NAME. Line 2 of the text display is blank. Enter a name that is not
already assigned to another library. Press CONFIRM to complete the library name entry.
Next, the pump displays ENABLE NONE LIBRARY OPTION with a Y response, indicating
that the option is currently enabled. Enabling this option adds a NONE selection to the list
of available libraries, allowing protocol selection to be bypassed during normal
programming. Press ▲ or ▼ if necessary to change the response and press CONFIRM to
complete this step. The pump displays UPDATING.
NOTE
All newly created libraries are enabled (available for selection at
programming time). A library may be disabled with the MODIFY
LIBRARY function.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
8.4.1.2.8 CHANGE CODE
The CHANGE CODE function allows the operator to change the library access code to any
other four digit number. The pump will subsequently allow access to the protocol library
editing mode only by use of the new library access code.
The pump displays <code> CODE where <code> is the current library access code. Press
CONFIRM to accept the current code, or enter a new code and press CONFIRM.
At the ENTER AGAIN prompt reenter the code from the previous step and press
CONFIRM. The pump displays UPDATING.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
8.4.1.2.9 RECEIVE
The RECEIVE function allows an operator to copy the entire configuration file including
protocol libraries and library access code from one pump directly to a second pump, thus
permitting replication of configuration information from a master pump to multiple receiving
pumps.