SYSTEM CONFIGURATION Chapter 8
8 - 9
changes were made to the protocol, the pump displays UPDATING
immediately.
The mode assigned to a protocol will be available when that protocol is
selected during normal programming, even if the mode is not enabled
in the MODES configuration group.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
If a mode other than ALL MODES, SINGLE DOSE, MANUAL SCHEDULE, AUTO
SCHEDULE or CUSTOM DILUTION is assigned to the protocol, the pump displays
additional prompts for any of the following that apply to the mode selected:
Concentration
Dose or Rate
At each prompt dashes are displayed indicating that no default value is defined. Use the
digits (0-9) and the decimal point (.) as necessary to enter a default value or leave dashes
displayed if no default value is required. Press CONFIRM to complete each entry.
Next, the pump displays ALLOW BOLUS with a Y response, indicating that bolus delivery
is currently allowed by the protocol. Press or if necessary to change the response
and press CONFIRM to complete this step.
Next, the pump displays a BOLUS SIZE prompt. Dashes are displayed indicating that no
default value is defined. Use the digits (0-9) and the decimal point (.) as necessary to
enter a default value or leave dashes displayed if no default value is required. Press
CONFIRM to complete the entry.
Next, the pump displays a BOLUS MAX RATE prompt for entry of a bolus delivery rate
limit. The units for bolus max rate are the same as the dose units (for example, mcg/min
for MCG/MIN mode, mL/hr for ML/HR mode, and so forth). Dashes are displayed
indicating that no default value is defined. Use the digits (0-9) and the decimal point (.) as
necessary to enter a default value or leave dashes displayed if no default value is required.
Press CONFIRM to complete the entry.
When all default settings have been confirmed, the pump displays UPDATING.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
8.4.1.2.7 CREATE LIBRARY

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals