Chapter 2 THEORY OF OPERATION
2 - 18
2.3.1.1.3 Aborting Calibration Review
Calibration review may be aborted at any time by turning off the pump.
2.3.1.1.4 LED Indications During Calibration Review
During Calibration Review, the LEDs indicate the following:
• The ALERT LED is on when the pump is vertical.
• The BATTERY LED is on when the plunger capture detector is detecting a syringe
plunger.
• The STANDBY LED is on when the flange sensor is detecting a syringe flange.
• The LOCK LED is on when the barrel sensor is detecting a syringe barrel.
2.3.1.2 Calibration Set Mode
Calibration Set mode provides the following calibration functions (in this sequence):
1. Calibrate Plunger Driver Position Sensor
2. Calibrate Syringe Recognition Sensor
3. Calibrate Pressure Sensor
4. Calibrate VB_SENSE
2.3.1.2.1 Activating Calibration Set Mode
To activate Calibration Set mode press the key sequence 5,0,8,9 following completion of
the POST.
NOTE
The pump sounds an error beep for each key. These tones are
provided to discourage unauthorized access and may be ignored.
Following the software version display, the pump proceeds with the plunger driver position
sensor calibration sequence.
2.3.1.2.2 Calibration Set Sequences
NOTE
If Calibration Set mode is entered following a CHARGE DEPLETED,
FUSE FAILURE, or POWER FAULT alarm, the pump will skip directly
to the VB_SENSE calibration step.