Chapter 2 THEORY OF OPERATION
2 - 28
2.3.6.7 CHARGE DEPLETED Alarm
If a CHARGE DEPLETED alarm occurs at power up, the pump will allow the operator to
access VB_SENSE calibration by entering ā5089ā. The pump will skip directly to the
VB_SENSE calibration step.
2.3.6.8 FINISH EDITING Alarm
The text display, FINISH EDITING, is displayed when additional information is required
during editing before operations are resumed.
2.3.6.9 LOCK Keypress During Occlusion Alarm
The LOCK key will be invalid during an occlusion alarm, since START must be pressed to
clear the alarm after the occlusion is cleared. The pump will sound a single (illegal
keypress) beep if LOCK is pressed during an occlusion alarm.
2.3.6.10 Clear Effective Rate if Dose is Altered
In Standby state, the pump will clear the effective rate display whenever the Dose field is
reset to dashes due to a syringe change or opened for editing and altered in any way.
When the Dose field is closed the effective rate will be recalculated and redisplayed.