FUNCTIONAL TESTS Chapter 4
4 - 9
6. Switch the pump “OFF.
4.2.10.5 “Fuse Failure” Alarm Test: (*D*)
1. Ensure that the pole clamp and battery are removed from the pump (refer to Chapter 7
for details). Ensure that the battery is connected to the pump.
2. Connect the AC charger to the pump and into an AC receptacle.
3. Switch the pump “ON” and allow the pump to complete the POST.
4. Disconnect the battery stack from the pump.
5. (*D*) After a short period of time (approximately a minute or so), the display will read
FUSE FAILURE, the ON CHARGE, BATTERY and SYSTEM LEDs will be on and the
pump will alarm continuously with alternating high-low beeps.
6. Switch the pump “OFF.
7. Reinstall the battery, battery cover and pole clamp, as necessary (refer to Chapter 7 for
details).