Chapter 5 ROUTINE MAINTENANCE
5 - 16
variable DC power supply to their “OFF” positions. If recalibration is required, check
“Fail” on the data sheet. Then reconnect the battery, enter access code
5089,
toggle to the desired calibration sequence, and recalibrate as previously specified.
20. Disconnect the power supply from the pump.
21. Reconnect the battery stack connector to the wiring harness ensuring that the tab is
well seated. Place the battery stack into the battery compartment and ensure that all
leads are properly dressed and are not crushed or kinked.
22. Carefully slide the cover onto the case. Use slight thumb pressure to ensure that the
cover seats properly.
23. Secure the cover to the case using the three 2-32 x 3/16 pan head screws being
careful not to overtighten them.
24. Sign and date the data sheet.