92 pro+ Surface Instruction for Use (209196 REV 8)
Cleaning and Disinfecting
c. Examine the following for damage:
•Top surface cover
•Bottom surface cover
•Zipper closure
d. Replace any damaged components.
STEP 2: Disinfection
a. With a new or clean wiping cloth soaked in an approved
cleaner/disinfectant, use light pressure to wipe all exterior
surfaces of the surface previously cleaned.
b. Make sure all surfaces remain wet with the
cleaner/disinfectant for the specified contact time. Re-wet
surfaces with a new wiping cloth as necessary. See
“Table 1:
Approved Cleaners/Disinfectants” on page 87 for the contact
time.
NOTE:
If bleach is used with another cleaner/disinfectant, use a new or clean
cloth/wipe soaked in tap water to remove any disinfectant residue
prior to and after the bleach application.
If applicable, clean the
x-ray sleeve.