36 pro+ Surface Instruction for Use (209196 REV 8)
VERSACARE Bed—pro+ Surface Installation Instructions
VERSACARE BED—PRO+ SURFACE INSTALLATION
INSTRUCTIONS
Tools: T10 Torx screwdriver
Parts: Long Power Cord Models
Short Power Cord Models
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—Make sure that hands, arms, legs, and feet are not
under the bed or between the sleep deck sections as they move.
• Warning—During this procedure, you will need to use an
alcohol-based cleaner. Such cleaners are flammable and toxic to
skin, eyes, and respiratory tract. Do not use the cleaner near an
open flame or in confined areas.
• Warning—Do not connect the power cord to an extension cord
or multiple outlet strip. There is a risk of overheating and fire may
cause injury or damage.
Any reference to a side of the bed or surface is from the patient’s view
lying in the bed on his or her back.
P3255A01
or
VersaCare pro+ surface with x-ray
P3255A02
or
VersaCare pro+ surface without x-ray
P3255A05 VersaCare pro+ MRS surface with x-ray,
Aus/NZ
P3255A03
or
VersaCare pro+ surface with x-ray
P3255A04 VersaCare pro+ surface without x-ray