86 pro+ Surface Instruction for Use (209196 REV 8)
Cleaning and Disinfecting
CAUTION:
To help prevent equipment damage, obey these cautions:
Caution—Do not steam clean or power wash the surface.
Pressure and excessive moisture can damage the protective
surface and its electrical components.
Caution—Do not use harsh cleansers/detergents, heavy duty
grease removers, solvents such as toluene, xylene, or acetone,
and do not use scouring pads (you may use a soft bristle brush).
Caution—Do not use bleach as your primary everyday
cleaner/disinfectant.
Caution—If applicable, fully extend the foot section prior to the
cleaning and disinfection process.
Caution—Do not launder any surface components.
RECOMMENDATIONS
For proper cleaning and disinfection, users should be trained.
The trainer should carefully read the instructions and follow them when
the trainee is being trained. The trainee should:
Be given time to read the instructions and to ask any questions.
Clean and disinfect the product while the trainer supervises.
During, and/or after this process, the trainer should correct the
trainee of any differences from the instructions for use.
The trainer should supervise the trainee until the trainee can clean and
disinfect the bed as instructed.
Baxter recommends to clean and disinfect the surface before first patient
use, between patient use, and regularly during extended patient stays.
Some fluids such as iodophor and zinc oxide creams, can cause
permanent stains or damage. Remove temporary stains by wiping
vigorously with a lightly-dampened wiping cloth.