pro+ Surface Instruction for Use (209196 REV 8) 89
Cleaning and Disinfecting
STEP 1: Cleaning
a. As necessary, first remove visible soil from the bed and the
surface using a wiping cloth soaked with an approved
cleaner/disinfectant (see “Table 1: Approved
Cleaners/Disinfectants” on page 87).
• Give special attention to seams and other areas where soil
may accumulate.
• A soft bristle brush may be used to loosen hardened soil.
• Use as many wiping cloths as needed to remove the soil.
It is important to remove all visible soil from all areas before
continuing to remove non-visible soil.
b. With a new wiping cloth soaked in an approved
cleaner/disinfectant, use firm pressure to wipe all surfaces of the
bed and pro+ surface. Use a new or clean wiping cloth as often as
necessary. Make sure the following items are cleaned (if
applicable):
d. If the integrated Centrella pro+
surface is removed from the bed,
put the connector cover on to the
cable connector.
•Power cord or surface
cable and sleeve
• Status indicator cord and
sleeve