pro+ Surface Instruction for Use (209196 REV 8) 89
Cleaning and Disinfecting
STEP 1: Cleaning
a. As necessary, first remove visible soil from the bed and the
surface using a wiping cloth soaked with an approved
cleaner/disinfectant (see “Table 1: Approved
Cleaners/Disinfectants” on page 87).
Give special attention to seams and other areas where soil
may accumulate.
A soft bristle brush may be used to loosen hardened soil.
Use as many wiping cloths as needed to remove the soil.
It is important to remove all visible soil from all areas before
continuing to remove non-visible soil.
b. With a new wiping cloth soaked in an approved
cleaner/disinfectant, use firm pressure to wipe all surfaces of the
bed and pro+ surface. Use a new or clean wiping cloth as often as
necessary. Make sure the following items are cleaned (if
applicable):
d. If the integrated Centrella pro+
surface is removed from the bed,
put the connector cover on to the
cable connector.
•Power cord or surface
cable and sleeve
Status indicator cord and
sleeve