pro+ Surface Instruction for Use (209196 REV 8) 31
CENTRELLA Smart+ Bed
1.2 CENTRELLA Smart+ Bed—Integrated pro+
Surface
Tools: T25 Tor x screwdriver
Parts:
If the Centrella Smart+ Bed needs to be upgraded to use the pro+
surface, refer to the upgrade instructions (210666).
P7923A01 pro+ Surface with x-ray, 36" (91 cm)
or
P7923A02 pro+ Surface with x-ray, 40" (102 cm)
or
P7923A03 pro+ Surface without x-ray, 36" (91 cm)
or
P7923A04 pro+ Surface without x-ray, 40" (102 cm)
Step 1: Setup
1. Make sure the brake is set on the bed.
2. Plug the bed into AC power.
3. Adjust the bed to the flat position.
4. Fully extend the foot section.
5. Put the bed in the lowest position.
6. Lower the siderails.
7. Remove the footboard.
8. If applicable, remove the surface (see the bed’s service manual).
Warning—Failure to unplug the bed could cause injury or
equipment damage.
9. Unplug the bed.